Price
$2,418.00 CAD
Stock
6 In Stock

  • Input Type - Toroidal (Electrodeless)
  • Programming Option - Advanced
  • Digital Communications Option - HART
  • Lightning Suppressor - Included
  • Housing Type - Powder-Coated Aluminum
  • Mounting Options - Wall (Side Mount)
  • Agency Approvals - ATEX / IECEx
  • Tag - Mylar

Specifications

Brand

Certifications & Approvals

  • CE
  • CSA
  • FM
  • ATEX
  • FISCO
  • IECEx
  • EMC

Environmental Protection

  • IP65
  • NEMA 4X

Temperature Element

  • Pt1000
  • 3 k Ω RTD
  • 4.75 k Ω RTD
  • Pt100

Temperature Display Range

  • -20° to 300° C (-4° to 572° F)

Cell Constant

  • 0.01, 0.10, 1.00

Measurement Range

  • 0 to 1,999 mS/cm
  • 0 to 19.99 mS/cm
  • 0 to 199.9 μS/cm
  • 0 to 1,999 μS/cm

Display Resolution

  • 0.001 μS/cm
  • 0.01 μS/cm
  • 0.1 μS/cm

Accuracy

  • Display: ± 0.5 % of measurement Range per decade (TB82EC and TC)
  • Display: ± 1.0 % of measurement Range per decade ((TB82TE)
  • Output: ± 0.02 mA
  • Temperature: 1° C or 1° F

Repeatability

  • Display: ± 0.5 % of measurement Range per decade (TB82EC and TC)
  • Display: ± 1.0 % of measurement Range per decade (TB82TE)
  • Output: ± 0.02 mA
  • Temperature: 1° C to 1° F

Stability

  • Display: ± 2 LSD typical, ± 5 LSD max
  • Output: ± 0.01 mA
  • Temperature: 1° C or 1° F

Operating Temperature

  • -20° to 60° C (-4° to 140° F)

Storage Temperature

  • -40° to 70° C (-40° to 158° F)

Humidity

  • Up to 95 % Relative Humidity

Enclosure / Body Material

  • Anodized Aluminum Alloy

Power Supply

  • Standard & HART: 14.0-53 Vdc (14.0-42 Vdc for certified applications)
  • FOUNDATION Fieldbus & PROFIBUS PA bus: 9-32 Vdc (non-I.S. model), 9-24 Vdc (I.S. model)

Mounting

  • Panel Mounting
  • Pipe Mounting
  • Wall Mounting

Documentation

show more