Price
$204.00
CAD
Stock
100+ In Stock
- Transmitter Type - Wall-/ Pipe-Mount (Gland Fittings)
- Voltage - 115 Vac
- Communication & IO - Additional Analog Output
- Sensor Type - Flow-Through System (ISO 7027 Compliant)
- Sensor Range - 0 to 40 NTU (without Auto-Clean)
- Second Dry Calibration - Not Included
- Sensor Cable Length - 10 m (32.8 ft)
- Manual - English
Specifications
Brand
Mounting
- Wall Mount
- Panel-mount
- Pipe-mount
Range
- High Range 0 to 400 NTU
- Low Range 0 to 40 NTU
Linearity
- Analyzer: ± 0.1 % FSD
- Sensor Max. < 1.0 %
Accuracy
- Display: ± 0.2 % of Reading, ± 1 Digit
- High Range: ± 5 % of Reading or 0.3 NTU
- Low Range: ± 2 % of Reading
Repeatability
- 200 to 400 NTU: 2 %
- 0 to 200 NTU: < 1 %
Response Time
- T90 < 1 min at 1 L/min-1 (0.26 gall [US]/min-1)
Flow Rate
- 0.5 to 1.5 L/min (0.13 to 0.39 gall [US]/min)
Limit of Detection
- High Range: 0.3 NTU
- Low Range: 0.003 NTU
Operating Temperature
- Analyzer: -20° to 55° C (-4° to 131° F)
- Sensor: 0° to 50° C (32° to 122° F)
Storage Temperature
- Analyzer: -25° to 55° C (-13° to 131° F)
Sample Pressure
- Up to 3 bar (43.5 psi)
Humidity
- Up to 95 % Relative Humidity, Non-Condensing
Material of Construction
- Cell Body unit: Black POM (Polyoxymethylene) Copolymer
- Wiper unit: EPDM
- Cell Body unit: Teflon
- Cell Body unit: Nickel-Plated Brass
- Cell Body unit: Polyamide 6
- Cell Body unit: Epoxy preform (Cured): Uni-forms 5034-00
- Cell Body unit: Spectrosil 2000 fused Silica
- O-Ring: Nitrile
- Wiper unit: 2-part Epoxy Adhesive (Cured)
- Wiper unit: Black Polycarbonate, 10 % glass fibre filled
- Wiper unit: Stainless Steel (SS 316 S13/S11) w/ Chemical Black
- Wiper unit: Silicone grease
Display
- 5-Digit x 7-segment backlit LCD
Power Supply
- 100-130 V, 200-260 V, 50/60 Hz
Power Consumption
- < 6 VA AC
Documentation
Brochure
- Information - 4690 Series Turbidity sensors 4670 to 4690 turbidity system upgrade pdf 241 KB
- Information - 4690 Series Turbidity sensors Emitter and receiver modules end cap replacement pdf 1.2 MB
- Information - 4690 Series Turbidity sensors Emitter and receiver modules replacement pdf 878 KB
- Information - 4690 Series Turbidity sensors Emitter module LED / lamp replacement pdf 1.2 MB
- Information - 4690 Series Turbidity sensors Feed / Drain assembly servicing pdf 659 KB
- Information - 4690 Series Turbidity sensors O-ring seal replacement pdf 936 KB
- Information - 4690 Series Turbidity sensors Wiper blade replacement pdf 1.1 MB
- Information - 4690 Series Turbidity sensors Wiper module upgrade pdf 1.2 MB
- Course description - Z253e Continuous Water Analyzer (CWA) Range | External version pdf 49 KB